Setting Bath Temperature - Heidolph Hei-VAP Value Operating Manual

Rotary evaporator
Hide thumbs Also See for Hei-VAP Value:
Table of Contents

Advertisement

Available languages

Available languages

Operation

Setting bath temperature

Possible temperature range:
Water bath application 20°C - 100°C
Oil bath application 20°C - 210°C
With a pre-set temperature above 75 °C to 80 °C we recommend oil or
polyethylen glycol as heating bath uid.
Set "Oil" as medium if you work with polyethylen glycol. Please follow
temperature speci cation provided by manufacturer.
Danger/ Danger
Risk of explosion!/ Danger d'explosion !
Potential loss of life/ Danger de mort
Warning / Avertissement:
Risk of scalding! / Risque de brûlures !
If you operate your device without a safety hood and at high heating
bath temperatures there is a risk of scalding./
En cas d exploitation de l unité sans capot de protection et à des
températures de chau age élevées, il existe un risque de brûlures.
Please wear safety clothes (lab coat, gloves, safety glasses) every time
you are in near proximity to the device during operation./
Porte des v tements de protection (blouse de laboratoire, gants,
lunettes de protection) à chaque fois que vous vous trouve à
proximité de l'unité lors de son exploitation.
Caution / Attention:
Potential damage of device / Dommages potentiels sur l'unité
The heating starts quickly and if empty, the heating bath can be
stained./
La chaleur augmente rapidement et si le bain marie est vide, il peut
être taché.
Switch on heating only when heating bath is lled su ciently with
uid./
Allumer le chau age uniquement lorsque le bain marie est
su samment rempli de liquide.
Only use
285 C ashpoint oils./
N utiliser que des huiles avec un point éclair
285 C.