Piping Omd-24 A; Figure 16 Piping Omd-24 A - Deckma Hamburg OMD-24 Series Instruction Manual

15 ppm bilge alarm
Table of Contents

Advertisement

7.4

Piping OMD-24 A

Connect the OMD-24 A Monitor to the sample point of the oily-water separator outlet
and to a source of oil free water. Fittings on the instrument are for 6 mm, 10 mm OD
copper or stainless-steel pipe. Adapter fittings for other pipe diameters are available upon
request.
The sample point should be located on a vertical section of the separator outflow piping
to minimize the effects of any entrained air. The tapping point should be at a level above
the outlet of the monitor to ensure the sample cell is flooded at all times.

Figure 16 Piping OMD-24 A