Rice Lake 150-10-6 Technical Manual

Rice Lake 150-10-6 Technical Manual

Digital physician scale waist-level

Advertisement

Quick Links

150-10-6
Digital Physician Scale Waist-Level
Software Version 11525
Technical Manual
August 18, 2022
PN 162342 Rev B

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 150-10-6 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Rice Lake 150-10-6

  • Page 1 150-10-6 Digital Physician Scale Waist-Level Software Version 11525 Technical Manual August 18, 2022 PN 162342 Rev B...
  • Page 2 All information contained within this publication is, to the best of our knowledge, complete and accurate at the time of publication. Rice Lake Weighing Systems reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
  • Page 3 Established a revision history; formatted content to match other medical manuals; software version 11525 Table i. Revision Letter History Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.ricelake.com/training or obtained by calling 715-234-9171 and asking for the training department.
  • Page 4: Table Of Contents

    7.0 Specifications ............... . . 23 Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost.
  • Page 5: Introduction

    Introduction Introduction The 150-10-6 Digital Physician Scale is efficiently designed to provide accurate, reliable and repeatable weight measurements. It is equipped with a handle and two heavy-duty rear wheels for easy portability. Manuals, resources and warranties are available from Rice Lake Weighing Systems website at www.ricelake.com/health...
  • Page 6: Safety

    Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.
  • Page 7: Assembly

    3. Gently thread the excess cable through the top of the scale column to eliminate any excess cable at the bottom of the column. This will eliminate the chance of pinching the cable as the column is being connected to the scale base. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 8: Wheel Assembly

    150-10-6 Digital Physician Scale 4. Attach the scale column to the base as shown in Figure 2-2 making sure not to pinch the cable. Figure 2-2. Attach Scale Column to Scale Base 5. Using the enclosed screws, attach the column to the scale base as shown in Figure 2-3.
  • Page 9: Attach Second Column To First Column

    The six AA batteries that come with the scale offer an average of 25 hours of continuous use. To install the batteries: 1. Open the battery chamber cover by loosening the thumb screw. 2. Insert batteries. Figure 2-7. Batteries in Battery Chamber 3. Close the battery chamber cover. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 10: Ac Power Connections

    Use the optional 120 VAC adapter or 230 VAC adapter when power is available. The optional AC power adapter plugs into the back of the indicator as shown in Figure 2-8. Rice Lake Weighing Systems offers optional AC adapters, utilizing an adapter not supplied by Rice Lake Weighing Systems voids all warranties. RS-232 Connection Connect the optional AC power adapter.
  • Page 11: Operation

    Pressing and holding Enter during startup will display ID. This is the first setup on entering into configuration mode. ENTER Up/Down Arrows Adjusts the value of the flashing digit/number. Adjusts height input (0.5 in/0.5 cm) while in BMI mode. Table 3-1. Scale Key Functions © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 12: Weighing

    150-10-6 Digital Physician Scale Weighing Use the following steps to weigh a patient. 1. Press to turn on the scale. 0.0 displays along with ZERO on the upper display. 2. Have the patient step on the scale. The patient’s weight is displayed, the LOCK annunciator is on and the indicator beeps to indicate the end of the weighing process.
  • Page 13: Troubleshooting

    Note: Load cell connection point is located underneath the curved plastic cover of the indicator. Remove four back retaining screws, remove curved back cover to access load cell connection point. Table 3-2. Troubleshooting © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 14: Configuration

    150-10-6 Digital Physician Scale Configuration Options and parameter setup are done through the scale configuration. Access to the setup switch is located inside the battery cover (shown below). Setup switch location Figure 4-1. Setup Switch Location General Navigation Use the buttons on the front panel to navigate through the menus and parameters.
  • Page 15: Configuration Mode

    Section 4.6 on page 17 Figure 4-2. Top Level Menu 6. Press to advance to the desired menu. 7. Press and advance in the manual to the related menu selection for further instructions. ENTER © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 16: Programming Mode Menu

    150-10-6 Digital Physician Scale Programming Mode Menu Various parameters can be set while in Programming mode. PROG FULL LOAD G-CAL G-LOC ASTRT ROUND DISP value value value value value value value 00000 0.0000 0.000 0.00 10.0 20.0 BAUD ATOL ALEN...
  • Page 17 0.00 2. Use the Up/Down arrow keys to change the available parameters. 3. When done, press ENTER key to move to the next parameter (BAUD). Table 4-1. Configuration Mode Menu (Continued) © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 18 150-10-6 Digital Physician Scale Parameter Description Choices Steps BAUD Baud rate Indicator display illustrates first two digits of baud rate only; The display toggles between a numeric value and baud; If you do not want to change this value, press the BMI key to move to the 1152 next setting;...
  • Page 19 OP7 - Bat type — (0=dry batteries, 1=rechargeable batteries) OP8 — OIML mode - (0=disable, 1=enable Press the BMI key to cycle back to the top level menu choices, ie: PROG/CALIB/DEF/DONE Table 4-1. Configuration Mode Menu (Continued) © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 20: Default Menu

    150-10-6 Digital Physician Scale Default Menu The default menu is used to return the scale back to its factory settings and is shown below. DONE PROG CALIB TEST Figure 4-4. Default Menu Use the following steps to return the settings back to their factory default.
  • Page 21: Test Menu

    Value Displays the current software version Value Displays the current battery level VALUE Value Displays the actual value Value Displays the actual raw counts of the scale Table 4-2. Test Menu © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 22: Communication

    The unit comes with an RS-232 port that enables weight data to be transmitted to other equipment, such as a computer or printer. The RS-232 cable with DB-9 connector (PN 100719) is available from Rice Lake Weighing Systems. That connection is shown in USB Connection section.
  • Page 23: Communication Protocols

    Example: <ESC><CUOM=m><ESC><E sets the unit of measurement Table 5-3. ESC Characters Name of Control Identifier Unit Unit of Measure (metric or constitutional) c (m or c) Table 5-4. Scale Global Values and Identifiers © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 24: Maintenance Protocol

    150-10-6 Digital Physician Scale Samples of Escape Protocol Examples of what is sent to the computer from the scale. direction of communication <SCALE> <PC> <SCALE> <PC> <ESC> R <ESC> W0210.9 <ESC> Nm <ESC> E <ESC>R <ESC>E Reading Reading Unit of Measure...
  • Page 25: Usb Connection

    5. With weight on the scale, press and hold the Print button on the indicator for three seconds. The patient’s weight is sent to the PC. Figure 5-5. Patient Weight Displayed © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 26: Maintenance

    150-10-6 Digital Physician Scale Maintenance The following section provides instructions for maintaining and cleaning the unit. Do not immerse the scale in cleaning or other liquid solutions. IMPORTANT Do not use Isopropyl alcohol or other solutions to clean the indicator display surface.
  • Page 27: Specifications

    14.7 in W x 14.7 in L x 3 in H (37.5 cm W x 37.5 cm L x 7.6 cm H) Overall Height 48.4 in (123 cm) Certifications and Approvals E113986 Complies with ANSI/AAMI ES60601-1:2005/A2:2010/(R)2012 CAN/CSA-C22.2 No. 60601-1:14 © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 28 150-10-6 Digital Physician Scale www.RiceLake.com Visit our website...
  • Page 30 © Rice Lake Weighing Systems Specifications subject to change without notice. 230 W. Coleman St. • Rice Lake, WI 54868 • USA U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319 August 18, 2022 PN 162342 Rev B...