Checking The Receiver Group - Keysight N5224B Series Service Manual

2-port and 4-port pna microwave network analyzers 900 hz/10 mhz - 43.5/50 ghz
Hide thumbs Also See for N5224B Series:
Table of Contents

Advertisement

Troubleshooting
Measurement System Troubleshooting

Checking the Receiver Group

Equipment Used for These Tests
Equipment Type
Spectrum analyzer
a. Must be capable of measuring a signal at 7.61 MHz (analyzers with DSP version 4.0), or 7.44 MHz
(analyzers with DSP version 5.0), and 1 GHz.
For all of the following checks, refer to the block diagrams at the end of this
chapter and to any of the following that are appropriate:
"2-Port Configuration, Option 200, S/N Prefix <6021" on page 6-31
"Bottom RF Cables, Standard 2-Port Configuration, Option 200, S/N Prefix
<6021" on page 6-34
"2-Port Configuration, Option 205, S/N Prefix <6021" on page 6-46
"Bottom RF Cables, 2-Port, Option 205, S/N Prefix <6021" on page 6-49
"2-Port Configuration, Option 219, S/N Prefix <6021" on page 6-68
"Bottom RF Cables, 2-Port, Option 219, S/N Prefix <6021" on page 6-71
"4-Port Configuration, Option 400, S/N Prefix <6021" on page 6-137
"Bottom RF Cables, Standard 4-Port Configuration, Option 400, S/N Prefix
<6021" on page 6-140
"Bottom Assemblies, 4-Port, Option 405, S/N Prefix <6021" on page 6-153
"Bottom RF Cables, 4-Port, Option 405, S/N Prefix <6021" on page 6-156
"4-Port Configuration, Option 419, S/N Prefix <6021" on page 6-180
"Bottom RF Cables, 4-Port, Option 419 (Ports 1 and 2), S/N Prefix <6021"
on page 6-183
"Bottom RF Cables, 4-Port, Option 419 (Ports 3 and 4), S/N Prefix <6021"
on page 6-186
"Bottom Assemblies, 4-Port, Option 420, S/N Prefix <6021" on page 6-192
"Bottom RF Cables, 4-Port, Option 420 (Ports 1 and 2), S/N Prefix <6021"
on page 6-195
"Bottom RF Cables, 4-Port, Option 420 (Ports 3 and 4), S/N Prefix <6021"
on page 6-198
Getting Ready to Test
Before checking the assemblies, you must open the analyzer.
Use an antistatic work surface and wrist strap to reduce the chance of
electrostatic discharge for all of the procedures in this chapter.
1. Turn off the analyzer power.
2. Unplug the power to the analyzer and disconnect all front and rear panel
connections except installed jumpers.
4-52
Model or
Alternate Model or Part
Part Number
Number
8565E
a
856xE
Keysight N5224B/25B Service Guide