Remove the water pump(21) electrical connection from the water filling solenoid valve (17) bobbin
•
socket.
Remove the water pump inlet and outlet hoses.
•
Remove the water pump mounting screws.
•
Mount the new water pump and make the hose and electrical connections carefully.
•
5.12. Replacing Safety Valve ( see Figure 8)
Remove the safety valve(13) inlet and outlet hoses by loosening their connection nuts.
•
Remove the connection adaptors on the safety valve inlet and outlet.
•
Connect the connection adaptors to the new safety valve inlet and outlet by putting some liquid
•
gasket.
Mount the new solenoid valve and make the hose connections carefully.
•
5.13. Replacing Water Level Sensor ( see figure-8)
Disconnect the water level sensor (5) main PCB (7) connection.
•
Remove the water tank (6) cover.
•
Remove the water level sensor tightening nut with a wrench.
•
Take the water level sensor out.
•
Mount the new water level sensor.
•
Replace the water tank cover and insulate it by putting liquid gasket.
•
Make the electrical connections according to the electrical circuit diagram ( Figure 3 or Figure 4)
•
5.14. Replacing Heater
Disconnect the cables which are connected to the heater terminals.
•
Take the shelf carrier (12) and the heater protection sheet (11) out by removing the shelf carrier
•
mounting nut in the sterilization chamber.
Remove the heater mounting nuts carefully from the chamber bottom with a wrench and take the
•
heater out from the sterilization chamber.
Mount the new heater to the connection threads by putting some liquid gasket to ensure that there
•
is no leakage.
Make the electrical connections.
•
5.15. Replacing Plastic Panel (see Figure 10)
Remove the display PCB (31), the printer and PC connection sockets, which are on the panel
•
sheet.
Remove the panel fixing screws and take the panel out.
•
Remove the plastic panel.
•
Clean the surface of the panel sheet with cleaners, alcohol etc.
•
Paste the new panel by meeting the place of the display board and the keys on it.
•