Introduction; Description; Intended Use - Saadat SHOOKA User Manual

Electroshock
Table of Contents

Advertisement

1 Introduction

1-1

Description

SHOOKA is an easy to use, battery operated, automated external defibrillator (AED).
SHOOKA automatically analyses patient's ECG signal and identifies sudden cardiac arrest
(SCA) and advises if the electrical shock is needed. SHOOKA provides audible and visual
prompts during operation which simplify system usage. SHOOKA is manufactured based on
CPR guideline of AHA revised in 2010.
SHOOKA provides an internal automated self-test and indicates the device is ready to use or
not. SHOOKA operates with disposable battery and detects low battery condition and produces
related prompts for battery exchanging.
SHOOKA is designed to operate in two different modes: Adult and child.
1-2

Intended use

SHOOKA is designed to use in clinical environments and public places and used by trained
user with appropriate technical and clinical trainings.
SHOOKA automated external defibrillator (AED) are indicated for use on victims of sudden
cardiac arrest who are:
• Unconsciousness
• Absence of breathing
• Absence of pulse and other sign of circulation
When the victim is less than 8 years old or less than 25 Kg weight (newborns and under 1 year
old are not included), use child pad. Do not delay therapy to determine exact age or weight. If
child pads are not available, apply adult pads in the position as shown for a child and set AED
on child mode and use the AED.
Contraindications
Never use the
SHOOKA Automated External Defibrillators (AED) if the victim
• Is responsive or conscious,
• Is breathing, or
• Has a detectable pulse or other sign of circulation.
SHOOKA is not intended for resuscitation of newborns and under 1-year-old patients.
Also, resuscitation of pregnant women should be done provided that it is under an expert
(including an obstetrician and neonatologist) supervision.