Saadat SHOOKA User Manual page 18

Electroshock
Table of Contents

Advertisement

Step
Action
Wait until AED determines
11
next step
Warning
➢ The defibrillator delivers up to 200 J of electrical energy. When discharging the AED,
do not touch the disposable pads.
➢ If a person is touching the patient, bed, or any other conductive material in contact
with the patient during defibrillation, the delivered energy may be partially discharged
through that person. Clear everyone away from contact with the patient, bed, and other
conductive material before shock delivery.
➢ During defibrillation, air pockets between the skin pads may cause skin burns for the
patent. Apply pads so that entire electrode adheres to skin. Do not reposition the
electrodes once applied. If the position must be changed, remove and replace with
new pads.
➢ Pads that are dried out or damaged may cause patient skin burns during defibrillation.
Do not use pads that have been removed from foil package for more than 24 hours.
Also, pads should not be in continuous contact with the patient's skin for more than
24 hours. Do not use electrodes beyond expiration date. Check that electrode adhesive
is intact and undamaged. Replace pads after 50 shocks.
➢ Defibrillation may cause implanted devices to malfunction. Place pads away from
implanted devices if possible. Check implanted device function after defibrillation, if
possible.
➢ Do not analyze in a moving vehicle. Motion artifact may affect the ECG signal
resulting in an incorrect shock or no shock advised message. Stop vehicle and stand
clear of the patient during analysis.
Result
Perform CPR
(Audio prompt will be repeated each 30
seconds)
CPR Remaining Time MINUTE: SECOND
After 30 Pushes, Give Two Breaths
(Visual prompt GIVE 2 RESCUE
BREATHS
ANALYZING HEART RHYTHM,
DON'T TOUCH PATIENT
(Audio prompt)
AED determine next action according to
analysis and specify continue with step 7 or 10
18