Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
IEC 61000-4-11
Voltage interruptions
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of test level.
T
Guidance and manufacturer's declaration – electromagnetic immunity
The SHOOKA is intended for use in the electromagnetic environment specified below. The customer or the user of
the SHOOKA should assures that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Proximity fields from RF
wireless
communications
equipment
IEC 61000-4-3
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications
equipment
Test frequency
Band
(MHz)
(MHz)
385
380- 390
450
430- 470
Enclosure
Patient coupling
Signal input/output
parts
Input a.c. power
Signal input/output
parts
Input a.c. power
Signal input/output
parts
Input a.c. power
Input a.c. power
Enclosure
Port
Input a.c. power
PATIENT coupling
Signal input/output
parts
ENCLOSURE
ENCLOSURE
a)
a)
Service
Modulation
TETRA 400
modulation
GMRS 460,
FRS 460
±8 kV contact
±2 kV, ± 4kV, ± 8kV,
± 15 kV air
N.A
N.A
N.A
N.A
N.A
N.A
N.A
N.A
N.A
30 A/m
50 Hz or 60 Hz
Compliance level
N.A
3 V/m, 80 MHz - 2,7 GHz,
80% AM at 1 kHz
Refer to the following table
(table 9 of EN 60601-1-2:
2015)
Maximum
b)
power
(W)
Pulse
b)
1.8
18 Hz
C)
FM
±5 KHz
2
deviation
1 KHz sine
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
Electromagnetic
environment –
guidance
IMMUNITY TEST
Distance
LEVEL
(m)
(V/m)
0.3
27
0.3
28
36