6 Maintenance
In this section maintenance procedure of the SHOOKA AED is described:
6-1
Physical and Accessories
• Verify expiration date of pads
• Verify expiration date of battery
• Verify pad's cable is not damaged
• Verify case has no physical problem (broken...)
• Verify additional accessories are available
6-2
SELF-Test
SHOOKA provides self-test in order to verify device is ready to use or not.
These tests verify the following features:
• Self-test include:
1- Battery charge
2- ECG circuitry
3- Defibrillator charge and discharge with low voltage (200V)
4- Button test
5- Pad detection
If SHOOKA passes each test successfully the status indicator shows (
Otherwise displays (
Problems and common errors of the device could be detected by troubleshooting
procedure.
and Troubleshooting
×
) symbol.
✓
) symbol.
22