6-3

Troubleshooting

In the following table, issues which could be happen to SHOOKA AED and relative actions
are described:
Issue
Status indicator shows X
BATTERY IS TOO LOW!!!
Device will shut down
automatically
Visual prompt
SELF-TEST FAILED
Visual prompt
INTERNAL ERROR
Voice prompt
CHECK DEFIBRILLATOR
PAD
Connect AED Cable to the
Device
Visual prompt
Table 6: Troubleshooting
Action
when AED is OFF:
Turn on the AED and wait until automated self-test is
completed. If status indicator still shows X turn off the
AED and replace the battery and repeat the procedure
again. If issue continues, contact service department.
when AED is ON:
Turn the device OFF, after that turn the AED on and wait
until automated self-test is completed. If status indicator
still shows X turn off AED and replace the battery and
repeat the procedure again. If issue continues, contact
service department.
Replace the AED battery with a new battery pack.
Device will perform self-test again. If visual prompt
continue, contact service department.
Turn off the AED and then turn it on and wait until
automated self-test is completed. If status indicator still
shows X , turn off AED and replace the battery and
repeat the procedure again. If issue continues, contact
service department.
Turn off the AED and replace the battery then press
ON/OFF button. Let device performs a self-test. If issue
continues, contact service department.
Check defibrillator pad is correctly connected to patient
and AED.
Replace PAD with a new one.
If issue continues, contact service department.
Check defibrillator paddle connector is correctly
connected to the socket.
If issue continues, contact service department.
23