Ventilation Requirements

To ensure safety requirements, the specifications and measurement accuracy of the
analyzer, you must keep environmental temperature within the specified range by
providing appropriate cooling clearance around the analyzer or, for the rackmount type, by
forcefully air-cooling inside the rack housing. For more information on environmental
temperature to satisfy the specifications and measurement accuracy of the analyzer, see
"Specifications And Supplemental Information" in the E5061A/E5062A User's Guide.
When the environmental temperature around the analyzer is kept within the temperature
range of the operating environment specification (See Section "Operating Environments"
on page 18), the analyzer conforms to the requirements of the safety standard.
Furthermore, under that temperature requirement, the analyzer still conforms to the
requirements of the safety standard even when placing the analyzer with cooling clearance
as follows:
Back
Sides
Figure 2-2
Ventilation space at the installation
Chapter 2
Requirements
≥ 180 mm
≥ 60 mm (both right and left)
Installation
Environmental Requirements
19