Vega WEIGHTRAC 31 Operating Instructions Manual page 57

Radiation-based sensor for mass flow detection
Hide thumbs Also See for WEIGHTRAC 31:
Table of Contents

Advertisement

WEIGHTRAC 31 • 4 ... 20 mA/HART - four-wire
Code
Cause
Text mes-
sage
F036
– Error during software
update
Faulty pro-
gram memory
F037
– Error in RAM
RAM hard-
ware error
F038
– Connection cable to the
Slave instrument interrupted
Slave signals
– Instrument not defined as
failure
Slave instrument
F040
– Sensor defective
Hardware
error
F041
– Error in the measured value
recording
Photomultipli-
er error
F045
– Error on the current output
Error on the
current output
F052
– Invalid parameter adjust-
ment
Faulty config-
uration
F053
– Adjustment of the input not
correct
Adjustment
data input not
correct
F066
– Adjustment not yet carried
out
Faulty adjust-
– Error during adjustment or
ment
when entering the lineariza-
tion table
F068
– Faulty instrument settings
Count rate too
high
F072
– Faulty instrument settings
Limit exceed-
ed
F080
– Instrument error
System error
F086
– Error in the Fieldbus com-
munication
Communica-
tion error
8 Diagnostics and service
Rectification
– Repeat software update
– Exchanging the electronics
– Restart instrument
– Exchanging the electronics
– Check the connection cable
to the Slave instrument
– Define instrument as Slave
– Restart instrument
– Exchanging the electronics
– Restart instrument
– Exchanging the electronics
– Check wiring of the current
output
– Exchanging the electronics
– Carry out a reset
– Correct adjustment of the
input
– Carrying out adjustment
– Carry out linearization
– Carry out a reset
– Carry out a reset
– Restart instrument
– Call our service
– Restart instrument
– Call our service
57