Vega WEIGHTRAC 31 Operating Instructions Manual page 20

Radiation-based sensor for mass flow detection
Hide thumbs Also See for WEIGHTRAC 31:
Table of Contents

Advertisement

4 Mounting
20
5. Insert the screws (8), each with one wedge lock washer (9),
through the clamp collars (11).
6. Place a wedge lock washer (9) from below on each screw (8) and
screw one nut (12) on each of the screws.
7. Align the clamp collars (11) and tighten the nuts (12) evenly with
8 Nm (5.9 lb ft).
8. Check if the sensor (13) is fastened correctly.
Mounting - Source container
Fig. 9: Mounting the source container on the measuring frame
1
Screws M10 x 40 (4 pcs.)
2
Wedge lock washer M10 Nordlock (8 pcs.)
5
Nut M10 (4 pcs.)
7
Source container (SHLD-1)
1. Place the closed and locked source container (7) from above onto
the measuring frame.
The source container is very heavy. Therefore use a suitable lifting
device. For this purpose the source container is equipped with a
suitable eye-bolt for a lifting hook, etc.
2. Align the source container (7) with the holes.
Make sure that the source container is placed in the correct direc-
tion on the crossbeam.
3. Insert the screws (1) with one wedge lock washer (2) through the
flange of the source container (7).
4. Place a wedge lock washer (2) from below on each screw (1) and
screw one nut (5) onto each screw.
5. Align the source container (7) and tighten the nuts (5) evenly with
45 Nm (33.2 lb ft).
WEIGHTRAC 31 • 4 ... 20 mA/HART - four-wire
7
1
2
5