Saturation; Color Temperature; Color Space - Barco MDSC-8532 User Manual

32” 4k uhd surgical display
Table of Contents

Advertisement

4.1.4 Saturation

To adjust the saturation level
1. Bring up the OSD main menu.
2. Navigate to the Picture menu.
3. Enter the Saturation submenu.
The command bar Saturation is highlighted.
4. Set the saturation level as desired and confirm.

4.1.5 Color temperature

About color temperature presets
The available color temperature presets for your display are:
5600K
6500K
7600K
9300K
Native
User
Factory calibration – White point:
The White Color points associated with the Color Temperature: 5600K, 6500K, 7600K or 9300K are
factory calibrated with a consequent reduction of the maximum luminance compared to Native Color
Temperature.
Only in case the User preset has been selected it is possible to get access to the color regulation
commands to adjust the gain of red, green and blue primary colors.
To select a color temperature preset
1. Bring up the OSD main menu.
2. Navigate to the Picture menu.
3. Enter the Color Temperature submenu.
4. Select one of the available color temperature presets and confirm.
If you selected the User option, a new menu will be displayed allowing you to manually adjust the
gain of red, green and blue.

4.1.6 Color space

About color space presets
The available color space presets for your display are:
Native (LCD primaries uncalibrated)
ITU709
BT.2020
(*)
DCI-P3
(*)
BT.2020 reproducible colors are within the limit of the LCD panel color gamut.
Factory calibration – Color space:
RGB primary calibration, according to the selected standard, is performed within the physical
limitation of the LCD panel used.
Advanced operation
R5914105 /01 MDSC-8532
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdsc-8532 sstpMdsc-8532 12gp

Table of Contents