Dicom Calibration; Reset Dicom Calibration; Dicom Error Threshold; Image Scaling - Barco Coronis Uniti User Manual

Table of Contents

Advertisement

Advanced operation
Warning:

4.22.4 DICOM calibration

About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
To start DICOM calibration:
1.
Bring up the OSD main menu.
2.
Navigate to the Configuration > Calibration > Embedded QA menu.
3.
Select DICOM calibration to start the calibration.
Warning:
restored.
Note: After calibration, the compliance check will start automatically.

4.22.5 Reset DICOM calibration

About reset DICOM calibration
It is possible to restore the original (factory default) DICOM curve.
To reset the DICOM calibration:
1.
Bring up the OSD main menu.
2.
Navigate to the Configuration > Calibration > Embedded QA menu.
3.
Enter the DICOM Preferences submenu.
4.
Select Reset DICOM Calibration to restore the original (not corrected) DICOM curve.

4.22.6 DICOM error threshold

About DICOM error threshold
The threshold to define the DICOM compliance can be modified in steps of 5% starting from 5 to 30%. When
the maximum deviation is not bigger than the selected threshold, the compliance check will be OK.
To set the DICOM error threshold:
1.
Bring up the OSD main menu.
2.
Navigate to the Configuration > Calibration > Embedded QA menu.
3.
Enter the DICOM Preferences submenu.
4.
Set Error Threshold as desired and confirm.

4.23 Image scaling

About image scaling
Enabling image scaling will copy each individual pixel to one or more adjacent pixels so that the size of the
displayed image will be a multiple of the original image source video input signal.
Image scaling is only possible when the resolution of your display's video input signal is less than or
equal to half the maximum resolution of the display.
38
K5902079 /11
Coronis Uniti
Pressing a key during the compliance check will abort the check.
Pressing a key during calibration will abort the calibration, previous values will be

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mdmc-12133

Table of Contents