The device is intended for use in the electromagnetic environment specified below. The operator of the
device should assure that the device is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD),
IEC 61000-4-2
Electrical fast
transient/burst,
IEC 61000-4-4
Surge,
IEC 61000-4-5
Voltage dips and
interruptions on
power supply input
lines
IEC 61000-4-11
Page | 64
Use of this device adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this device
and the other equipment should be observed to verify that they are operating
normally.
The use of accessories other than those specified for the device is prohibited. They
may result in increased electromagnetic emissions or decreased electromagnetic
immunity of the device.
IEC 60601 Level Compliance Level
8 kV contact
8 kV contact
15 kV air
15 kV air
2 kV for power
supply lines
2 kV for power
1 kV for SIP/SOP
supply lines
lines
1 kV line to line
1 kV line to line
2 kV line to earth
2 kV line to earth
0% U
for 0.5
0% U
T
cycle
cycle
0% U
for 1 cycle
0% U
T
70% U
for 25/30
70% U
T
cycles
cycles
0% U
for 250/
0% U
T
300 cycles
cycles
Novoxel.com
 
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
for 0.5
T
Mains power quality should be that of a
typical commercial or hospital
for 1 cycle
environment. If the user of the equipment
T
requires continued operation during
for 25/30
power mains interruptions, it is
T
recommended that the equipment be
powered from an uninterruptible power
for 250/300
supply or a battery.
T
User Manual TXLD0012
Electromagnetic Environment -
Guidance
P/N T2-UM-E-ENG Rev 10