Chapter 5: Troubleshooting
The following Table describes all notifications/error screens which may be displayed during the operation
of the Tixel2 device:
Item #
The Handpiece which was pre-selected in the Home screen, will be heated to working
temperature (400°C) and sterilized. When heated the icon will become orange and blink.
The selected Handpiece shall be indicated by the red dot in the following icon:
When Handpiece temperature have reached 385°C and sterilized, the icon will become
green, and blink until treatment parameters are selected (see indents 4, 5 & 6 further below
in this table).
After the treatment parameters have been selected, the Handpiece will become active; The
green icon will be displayed continuously and treatment can be performed.
Note: The other Handpiece will remain at hibernation temperature (200°C) indicated by
an orange icon.
1.1
Preparing the Second Handpiece for Operation
Press on the hibernated Handpiece icon. The second Handpiece will be heated to working
1.2
temperature and sterilized for 1:00 minute. During the heating process, the icon will
become orange and blink.
After the sterilization process has been completed, the second Handpiece will have a
blinking green icon but without the red dot.
Switching between Handpieces
Press on the icon of the second Handpiece. The Handpiece icon will have a red dot, as
demonstrated below:
The icon will blink until treatment parameters are selected (see indents 4, 5 & 6 further
below in this table). After the treatment parameters have been selected, the Handpiece will
become active; The green icon will be displayed continuously and treatment can be performed.
"X" stands for either 1) signifying the left Handpiece or 2) signifying the right Handpiece.
P/N T2-UM-E-ENG Rev 10
 
Description
Novoxel.com
User Manual TXLD0012
Page | 35