Operating State, Location and Position

Do not place the product on surfaces or tables that (for reasons of weight or stability, etc.) are
unsuitable for this purpose.
Do not place the product on a heat-generating surface or near heat emitting devices such as heaters.
Verify that there is sufficient clearance between the product and any other system that may exhaust
heat.
The product is for indoor use only.
General Operating Conditions:
The product's operating ambient range is temperature of 10 to 30°C (50 to 86 °F) and relative humidity of
®
20-80%. Novoxel
recommends that an ambient temperature of 20 to 25°C (68 to 77°F) and relative
humidity of 40-60% be maintained during normal operation, as this will result in better performance and
in longer MTBF of the equipment. The temperature must not exceed the maximum temperature specified
above.
Maximum operating altitude: 1000 m ASL (~3,300 ft.).
Maximum transport altitude: 4500 m ASL (~15,000 ft.).

Electrical Safety Information

Electrical Medical Equipment Classification

Equipment Classification:
Equipment Type:
Safety Fuse:
P/N T2-UM-E-ENG Rev 10
 
The product may be operated only in full compliance with the operating
conditions and in the positions specified by Novoxel
The product's ventilation should not be obstructed.
Following are the required operating position and conditions:
Class I
B
Ceramic, T5A, 5×20 mm, Littelfuse Inc.