Note: When both Handpieces are active (indicated by a green icon), switching
between them may be performed instantly. The non-active Handpiece will be
indicated by a blinking green icon.
Automatic Hibernation/End of Session
The active Handpiece/Handpieces will remain heated for 20 minutes after the last
1.1
pulse has been performed. After this time, it/they will be automatically cooled down
1.2
to hibernation temperature (200°C).
After 10 further idle minutes, the treatment session will be automatically stopped and
(Continued)
the Home screen will be displayed.
Notes:
2
Indication of the selected Handpiece.
3
End of session and return to the Home screen.
Selection/De-selection of the "Double Pulse" option. This option will be displayed
4
for pulse durations of 8-16 milliseconds only. When selected, the icon will become
orange.
Pulse duration. Varies between 5-18 milliseconds. The increase/decrease is by
intervals of 2 milliseconds, except for a 1 millisecond gap between the 5-6 options.
5
6
Protrusion. Varies between 100-1,000 microns, in 100 micron intervals.
Temperature indication, cord to the next color index:
Ready to work: Red 385-400°; Orange 200-384°; Yellow 80-199°; Blue 0-79°.
7.1
Handpiece heating will be indicated by a small upward arrow icon (˄) located above
7.2
the Handpiece temperature indication. Handpiece cooling will be indicated by a
small downward arrow (˅).
Page | 30
 
Any activation after hibernation will require re-sterilization.
Any activation after end of session will require re-sterilization, unless the
Handpieces remain in working temperature.
Treat only after establishing adequate parameters. Not every
parameter setting is suitable for every patient. See clinical
information and guidelines for further information.
Novoxel.com
User Manual TXLD0012
P/N T2-UM-E-ENG Rev 10