The Handpiece which was pre-selected in the Home screen, will be heated to
working temperature (400°C) and sterilized. When heated the icon will become
orange and blink.
The selected Handpiece shall be indicated by the red dot in the following icon:
When Handpiece temperature have reached 385°C and sterilized, the icon will
become green, and blink until treatment parameters are selected (see indents 4, 5 &
6 further below in this table).
After the treatment parameters have been selected, the Handpiece will become
active; The green icon will be displayed continuously and treatment can be
performed.
Note: The other Handpiece will remain at hibernation temperature (200°C) indicated
by an orange icon.
1.1
Preparing the Second Handpiece for Operation
Press on the hibernated Handpiece icon. The second Handpiece will be heated to
1.2
working temperature and sterilized for 1:00 minute. During the heating process, the
icon will become orange and blink.
After the sterilization process has been completed, the second Handpiece will have a
blinking green icon but without the red dot.
Switching between Handpieces
Press on the icon of the second Handpiece. The Handpiece icon will have a red dot,
as demonstrated below:
The icon will blink until treatment parameters are selected (see indents 4, 5 & 6
further below in this table). After the treatment parameters have been selected, the
Handpiece will become active; The green icon will be displayed continuously and
treatment can be performed.