Two Handpiece configurations exist: Standard and Periorbital. These are in fact identical Handpiece
designs, which differ only by: Tip configuration and Distance Gauge configuration (see further
information below regarding the Distance Gauge). Each HP also automatically sterilizes the Tip before
each treatment session.
Both Handpieces are shown below:
The Handpiece consists of:
A trigger activated electric motor.
A motion control apparatus.
Tip heating element.
A Distance Gauge (or DG):
The Distance Gauge is attached to the Handpiece on its distal end by magnets. The tip is pushed forward
and backwards through it during the treatment pulse. The DG is placed on the patient's skin at the desired
location for treatment prior to a pulse. The Distance Gauge also functions as a safety precaution by
securing the operator and the patient from exposure to the hot Tip and to the heating unit. The DG can be
manually removed for cleaning of the Tip and for cleaning of the DG itself.
P/N T2-UM-E-ENG Rev 10
 
Distance Gauge
The Tixel2 Handpiece Configurations - Left is periorbital, Right is Standard
Trigger
Novoxel.com
User Manual TXLD0012
Page | 19