Annual Inspection by Certified Local Representative

Prior to activating the Handpiece, always check your surroundings to ensure safe
operation of the device. Always point the Handpiece away from yourself when
inspecting its action and integrity.
Visually inspect the system's integrity. Verify that all the system components are well secured intact.
In case the components are damaged, the device should not be operated and delivered immediately to
a certified service provider for servicing.
Visually inspect the power cable for any damage.
Inspect the Tip for cleanliness. In case the Tip is not clean, perform the cleaning procedure as
described in this document.
Activate the system, to verify that the software functions normally.
Set system parameters to 6ms × 600 µm. Activate the Handpiece by pressing the trigger to produce a
pulse and visually inspect the Tip's movement along its travel route forth and back. The movement
should be smooth without any substantial mechanical sound.
Repeat the test at 16ms × 800 µm.
Inspect the Tip's orientation within the Handpiece. Activate the system. Verify that the Tip moves
freely inside the lumen of the Distance Gauge during a pulse. The Tip should be centered. Slight
deviation from the DG lumen's center is allowed.
Perform a pulse to inspect the centering of the Tip throughout the forward and back motion.
P/N T2-UM-E-ENG Rev 10
 
Adequate Centering of the Tip within the Distance Gauge.
Novoxel.com
User Manual TXLD0012
Page | 61