Failure
Code
Text message
F013
No valid measured val-
ue available
F017
Adjustment span too
small
F025
Error in the lineariza-
tion table
F036
no operable sensor
software
F040
Error in the electronics
F041
Communication error
F042
Communication error
Secondary sensor
F080
General software error
F105
Measured value is de-
termined
F113
Communication error
F260
Error in the calibration
F261
Error in the instrument
settings
VEGABAR 87 • 4 ... 20 mA/HART
maintenance for the instrument because a failure is expected in the
near future (e.g. due to buildup).
This status message is inactive by default.
Cause
Gauge pressure or low pressure
Measuring cell defective
Adjustment not within specifi-
cation
Index markers are not continu-
ously rising, for example illogical
value pairs
Failed or interrupted software
update
Hardware defect
No connection to the sensor
electronics
No connection to the Second-
ary sensor
General software error
The instrument is still in the
switch-on phase, the measured
value could not yet be deter-
mined
Error in the internal instrument
communication
Error in the calibration carried
out in the factory
Error in the EEPROM
Error during setup
Error when carrying out a reset
9 Diagnosis, asset management and service
Rectification
Exchange measuring cell
Send instrument for repair
Change the adjustment accord-
ing to the limit values
Check linearization table
Delete table/Create new
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
Exchanging the electronics
Send instrument for repair
Check connection between sen-
sor and main electronics (with
separate version)
Check connection between Pri-
mary and Secondary sensor
Disconnect operating voltage
briefly
Wait for the end of the switch-
on phase
Disconnect operating voltage
briefly
Send instrument for repair
Exchanging the electronics
Send instrument for repair
Repeat setup
Repeat reset
DevSpec
State in CMD 48
Byte 5, Bit 0 of
Byte 0 ... 5
Byte 5, Bit 1 of
Byte 0 ... 5
Byte 5, Bit 2 of
Byte 0 ... 5
Byte 5, Bit 3 of
Byte 0 ... 5
Byte 5, Bit 4 of
Byte 0 ... 5
-
-
Byte 5, Bit 5 of
Byte 0 ... 5
Byte 5, Bit 6 of
Byte 0 ... 5
Byte 4, Bit 4 of
Byte 0 ... 5
Byte 4, Bit 0 of
Byte 0 ... 5
Byte 4, Bit 1 of
Byte 0 ... 5
53