4 Mounting
Filter element - Position
Filter element - Position
Ex-d version
16
6
1
6
4
Fig. 7: Position of the filter element
1
Plastic, stainless steel single chamber (precision casting)
2
Aluminium - single chamber
3
Stainless steel single chamber (electropolished)
4
Plastic double chamber
5
Aluminium, stainless steel double chamber housing (precision casting)
6
Filter element
With the following instruments a blind plug is installed instead of the
filter element:
Instruments in protection IP66/IP68 (1 bar) - ventilation via capillar-
ies in non-detachable cable
Instruments with absolute pressure
Turn the metal ring in such a way that the filter element points
downward after installation of the instrument. This provides better
protection against buildup.