Safety
INVU by Nuvo™ Usage Restrictions
2.6
INVU by Nuvo™ Usage Restrictions
INVU Sensor Band™ must be removed prior to:
Magnetic Resonance Imaging (MRI) scans
Computer Tomography (CT) scans
Direct x-rays
Defibrillator use
ECG and EMG connection
Any other potential exposure to a strong electromagnetic field.
2.7

Regulatory Classification

The INVU by Nuvo™ system, including INVU™ App, is classified as Class II (US).
2.8

Complying with Standards

The INVU Sensor Band™ was approved as per IEC 60601-1-2:2014, which contains tests of
immunity to proximity RF fields of different RF wireless communication equipment. The test
was implemented in accordance with clause 8.1 which defined the maximum power of the
wireless equipment and its minimum distance to the INVU Sensor Band™ at which the INVU
Sensor Band™ functionality stays steady. Thus, Portable RF communications equipment
(including peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the INVU Sensor Band™. Otherwise, degradation
of the performance of the INVU Sensor Band™ could result.
Note: Disposal of this device should be performed in accordance with local regulations and
WEEE (Waste of Electrical and Electronic Equipment). Alternatively, the user can send the
device for disposal back to the manufacturer or its local distributor.