Section 2
Unpacking and Installation

Introduction

2-2
Magnetic Stirrer
Environmental Conditions
The unit should be installed in a suitable environment as described below.
Avoid direct sunlight.
Room temperature should be 5°C ~ 40°C
Relative Humidity (RH%) should be less than 80%.
Altitude should be less than 2,000m.
Location Conditions
Place the unit on the bench or table when in use, and keep a proper
distance between it and other apparatuses.
• Do not use the plastic beaker when operating the unit in explosive
atmospheres.
• The plastic beaker can cause the magnetic field, resulting in igniting to
explosive substances.
• Do not place the unit near other laboratory equipment which can be readily
influenced, even by weak motor vibration, or by magnetic force because the
unit create a magnetic field.
Arrangement of Unit
Place the unit on the table. Check to make sure the non-slip feet are
secured tightly to the unit.
Ensure that the unit will be parallel to the table.
Setting Up Cover Plate on Top Plate
Cover silicon pad on top of the top plate for the safety test.
Silicon pad cannot be used according to user's option.