Fault – Cause – Remedy
Fault – Cause – Remedy
Fault
Display PROBE
Display BATT
No display for SpO
, pulse rate or
2
pulse strength
or
pulse rate is incorrect or fluctuates
or
SpO
is incorrect or fluctuates.
2
Oxipac 2000 does not switch on
when å key is pressed.
Oxipac 2000 switches off
suddenly.
Green
LED on charger does
not light.
No print-out from optional printer.
No print-out or meaningless string
of characters.
This "Instructions for Use" manual is published for public use and is only meant for your information.
We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
32
Cause
Sensor not correctly connected to
Oxipac 2000.
Sensor not properly attached to
patient.
Adhesive plaster fixing sensor too tight.
Faulty sensor or cable or Oxipac 2000.
Battery low
Patient is very restless.
Sensor not properly connected to
Oxipac 2000.
Sensor not properly attached to
patient.
Adhesive plaster fixing sensor too tight.
Faulty sensor or cable or
Oxipac 2000.
Battery discharged.
Oxipac 2000 faulty.
Battery discharged.
No mains power supply to charger.
No mains power supply at power
point.
Faulty charger or Oxipac 2000.
Printer not connected to mains power
or printer not switched on.
Connecting cable not properly
connected.
Transmission format on printer set
incorrectly, see page 20.
Faulty PIN configuration of D25-cable.
Printer faulty.
Remedy
Connect sensor plug correctly and
firmly to socket on Oxipac 2000.
Attach sensor properly, page 7ff.
Loosen adhesive plaster.
Call DrägerService.
Charge battery, page 24.
Keep patient from moving so much.
Set longer averaging interval, page 25.
Check all connections between sensor
and Oxipac 2000 for tightness.
Attach sensor properly, page 7ff.
Loosen adhesive plaster.
Call DrägerService.
Charge battery, page 24.
Call DrägerService.
Charge battery, page 24.
Check that mains plug on mains cable
is connected to power point and to
Oxipac 2000.
Check power point, e.g. with a lamp.
Call DrägerService.
Connect mains plug to power point
and switch on printer.
Check plug connection.
Adjust transmission parameters on
printer according to printer's
Instructions for Use.
Make sure PIN-configuration, is
correct, see section on PIN
configuration of cable, page 36.
Have printer repaired.