Fault, Cause, Remedy
Main module
Fault, Cause, Remedy
Main module
Fault
Cause
Remedy
Red
Alarm
LED is lit.
Alarm tone commences.
Fault in module.
Identify faulty module.
See "Fault, Cause, Remedy"
of module.
Red
Inop.
LED is lit.
Continuous tone commences.
Basic fault in electronics.
Switch incubator off and on again.
If
Inop
message not repeated:
press (X) key and reset values.
If
Inop
message repeated,
incubator not working.
Call DragerService.
Red
Inop.
LED is lit.
Continous tone commences,
also
Err
display in a module (except
air temperature control module).
Fault in specific module:
Skin temperature control module/
humidity control module.
Switch off specific module. All
other functions will remain
operational.
Call DragerService.
Red
N
LED is lit.
Continuous tone commences.
No mains supply.
Fault in incubator.
Check that mains plug is
connected to power supply.
Check that power is being supplied.
Inform in-house technical staff.
Call DragerService.
38