Acceptance and handover

Delivery
Inspect delivery
Check the shipping crate for
damage Ensure that the delivery
is complete: Instructions for Use /
Installation Instructions
Caution!
Prior to initial operation of terminal units and after exchange of compressed gas tubes in medical
supply units, the following tests are to be carried out in accordance with DIN EN ISO 7396-1
-
Leakage test;
-
Blockage test;
-
Particle pollution test;
-
Testing the gas-type.
If these tests do not pass inspection, the medical devices are not allowed to be put into Operation.

Troubleshooting

Fault – Cause – Remedy
WARNING
Repair and maintenance of the medical device
must be carried out by trained technical
personnel.
Dräger recommends that a service contract is
obtained with DrägerService and that all
repairs are performed by DrägerService.
Dräger further recommends that only authentic
Dräger repair parts are used for maintenance.
Otherwise, the correct functioning of the medical
device may be compromised.
See "Maintenance" chapter.
Fault
Insufficient gas supply,
hissing sound.
Cause
Pressurized gas is escaping from
inside the supply unit.
Final Test
(acc. to acceptance protocol)
The acceptance test includes:
Gas-type inspection
Inspection of electrical circuits (also for
emergency power outlets, if applicable)
Ground continuity test
Leakage current test
Remedy
Notify the hospital maintenance
or DrägerService to check the
gas supply. Take care of patients
and connected consuming
devices.