3.11.2 Program the process input

• Go to menu C1.10 via "Setup
process input
C1.10
standard
flow mode
flow mode
flow mode
flow mode
cold
or
process input
C1.10
standard
flow mode
flow mode
flow mode
flow mode
heat
heat
heat
heat
• Choose "Heat" or "Cold" to activate energy measurement.
process input
C1.14
A: supply B: return
current inputs
current inputs
current inputs
current inputs
A: supply B: return
A: supply B: return
A: supply B: return
A: supply B: return
• Choose in "current inputs
INFORMATION!
In case the function is "Heating", the temperature at the "supply" side is the highest.
In case the function is "Cooling", the temperature at the "supply" side is the lowest.
10/2020 - 4006193902 - HB OPTISONIC 6300 V2 R02 en
Process input
flow mode
" which sensor is located at the supply side of the process.