Regenerant Line Connections To And From The Regenerant Reservoir - Thermo Scientific Dionex ACRS-ICE 500 Product Manual

Suppressor
Hide thumbs Also See for Dionex ACRS-ICE 500:
Table of Contents

Advertisement

2
– Installation

2.4 Regenerant Line Connections to and from the Regenerant Reservoir

!
Refer to the pressurizable plastic reservoir manual (Document No. 032677) for complete
installation instructions.
NOTE
A. Install the Dionex ACRS-ICE 500 in the slot inside the chromatography module. Use a
B. Using a coupler (P/N 034056) and a 1/8" o.d. piece of tubing that comes in the
C. Install a restrictor to the regenerant flow by connecting a piece of 0.012" i.d. tubing to
D. Install the regulator (P/N 038201) on the helium or nitrogen supply line. The
E. Prepare the regenerant. Fill the regenerant reservoir. Make sure that the O-ring is inside
F. Adjust the reservoir pressure (0–25 psi) to deliver a regenerant flow rate of 2–3
!
A safety relief valve on the reservoir regulator prevents pressure greater than 25 psi from
being applied to the regenerant reservoir.
NOTE
Thermo Scientific
032661-12
coupler (P/N 039056) to connect one end of the 30" tubing assembly (P/N 035727) that
comes in the Installation Kit to the regenerant reservoir. Push a slip-on filter (P/N
038260) on the end of the line that goes inside the reservoir. Connect the other end of
this tubing to the REGEN IN port of the Dionex ACRS-ICE 500 (see Figure 1).
Installation Kit, connect one end of this line to the REGEN OUT port of the Dionex
ACRS-ICE 500 (see Figure 1). Direct the other end of the line to waste.
the end of the waste line that is approximately 12" long (see Figure 3).
regenerant flow rate is regulated by two factors: the pressure applied to the regenerant
reservoir and the length of the 0.012" i.d. tubing connected to the waste line.
the cap of the reservoir before screwing the cap onto the reservoir. Screw the cap onto
the reservoir tightly and place the reservoir near the chromatography module. Refer to
Section 3.2.