2. The content of the next screen displayed depends on the Processor Assembly type. To
determine your instrument's Processor Assembly type, see
Processor Assembly Type" on page
If the instrument has a PC8 or PC9 Processor Assembly, then the screen contains the
selections shown in
Figure 9
Alignment Data Wizard Storage Options for PC8/PC9
The available storage location options are as follows:
— Internal Flash is the Calibration Backup memory in the A4 Processor assembly. For
PC8/PC9, this is the default location.
— External Removable Drive. This is the SD memory card in the A4 Processor assembly. This
could also be a manually selected drive or an external USB drive. Use the browse feature to
select the actual destination as needed.
When naming the file, it is recommended to use the instrument model number, serial number,
and date to help identify the calibration data file with the correct instrument.