Environmental Requirements; Biosafety - PerkinElmer OPERA PHENIX PLUS Manual

Site readiness
Table of Contents

Advertisement

Country Kit
HH14001610
(JP)
HH14001611
(Export)
HH14001612
(CH)
HH14001613
(Nordics)
HH14001614
(East Europe)
2.5

Environmental Requirements

Operating conditions:
l
o
o
Storage temperature: 10 °C to 40 °C (50 °F to 104 °F)
l
Indoor use only
l
A climate-controlled, clean and dust-free laboratory is recommended
l
Atmospheric pressure up to 2000 m altitude
l
Vibration-free work surface
l
2.6

Biosafety

Opera Phenix Plus – Site Readiness
Used for
Keyboard
layout
Japan
US QWERTY
European
UK QWERTY
Export
Switzerland,
DE QWERTZ,
Liechtenstein
AZERTY
Sweden,
US QWERTY
Finland,
Denmark
Poland,
US QWERTY
Hungary
+20 °C to +25 °C (68 °F to 77 °F)
Relative humidity < 70 %, non-condensing
Danger! Biohazard!
This device is designed for analyzing biological substances at
biosafety level 1 and 2 (BSL-1 or BSL-2).
Do not use more hazardous substances due to limited
l
possibilities of decontamination.
Observe the safety instructions for the substances used.
l
Observe the instructions for decontamination and cleaning.
l
Always make sure to comply with national and international
l
laboratory safety instructions.
Always wear your personal safety equipment.
l
2 Requirements
Power plug
Plug type/norm
JIS C 8303
498 GJ
KC-015A
IEC 60884-1
SEV 1011
SEV 6534/2
KC-015A
IEC 60884-1
KC-015A
IEC 60884-1
21