INTELLIDOX DOCKING MODULE
MENU
5.
Configure the gas type for the blend.
a)
To configure a gas type in a multi-gas blend, select Gas blend from the Inlet
edit menu, and then press
the blend.
Select gas type from the Inlet edit menu, and then press
b)
select the type of gas. Verify that the gas type selected matches the type of
gas contained in the calibration cylinder.
c)
Press
to abandon the changes and return to the Inlet edit menu.
6.
Configure the gas concentration.
a)
Verify that the gas concentration displayed on the screen matches the gas
concentration contained in the calibration gas cylinder. If the concentration
does not match, select Concentration: from the Inlet edit menu, and then
press
b)
Press
c)
Press
to abandon the changes and return to the Inlet edit menu.
7.
Record the lot number of the calibration gas cylinder.
a)
Select the lot number setting from the Inlet edit menu.
b)
Press
alphabetic, numeric and special characters. Press
next character space. Up to 14 characters may be entered.
c)
Press
to abandon the changes and return to the Inlet edit menu.
8.
Repeat steps 3, 4, 5, and 7 until all of the gas inlets are configured.
to close the Inlet edit menu and return to the Inlet overview. Press
9.
Press
again to return to Adjust dock station settings.
HONEYWELL
USER MANUAL || DOCK STATION SETTINGS
or
to select one of the gases available in
to save the changes and return to the Inlet edit menu, or press
to edit.
or
to configure the gas concentration.
to save the changes and return to the Inlet edit menu, or press
to edit the lot number, and then press
to save the changes and return to the Inlet edit menu, or press
or
to find and
or
to find and select
or
to move to the
PAGE 42 OF 119