INTELLIDOX DOCKING MODULE

Calibration Equipment and Gases

To receive a complete list of calibration gases and accessories approved for use with this
product, contact Honeywell or an authorized distributor, or visit the Order Information
section of the product website at sps.honeywell.com.
For BW Ultra sticky gases CL
Ammonia (NH
between calibrations and all the nests used in gang should be for BW Ultra. If it is not the
case, it is highly recommended to perform calibration for NH
reason, when NH
Other sticky gases not mentioned here are not available to be calibrated in BW Ultra using
IntelliDox.
Other products are not set to calibrate sticky gases but BW Solo that includes NH3, CL2,
NO2, HCL and HF.
HONEYWELL
The use of calibration gas cylinders other than those specified by Honeywell may
result in unsafe calibration or possible non-recoverable failure of the equipment, and
will invalidate the warranty.
A demand flow regulator must be used with all calibration gas cylinder connections.
The module inlet should not be pressurized.
2
3
) the calibration is available. However, it requires a wait period of 15 minutes
3
, Ammonia is set on bays 2 to 5 a warning will be displayed as .
USER MANUAL || MAINTENANCE
, NO
2
, and HCN calibration is available in all bays. For
3
only on the first bay. For this