Digital Remote Controlled Table with dynamic flat panel detector
FAILURE MESSAGES (TO BE CONTINUED)
CARRIAGE EOT +
CARRIAGE EOT -
COLUMN EOT +
COLUMN EOT -
TILTING EOT +
TILTING EOT -
TABLE HEIGHT EOT +
TABLE HEIGHT EOT -
LONG TABLE TOP EOT +
LONG TABLE TOP EOT -
S.I.D. EOT +
S.I.D. EOT -
LAT TABLE TOP EOT +
LAT TABLE TOP EOT -
BACK TO AUTOPOSITIONING
MANUAL COLLIMATOR
Stitching not Possible, Move Carriage
Tomography not Possible, Move Carriage
User's Manual
This indicates a system failure with the carriage end position on foot side and
forbids all the movements
This indicates a system failure with the carriage end position on head side and
forbids all the movements
This indicates a system failure with the column incidence end position on foot
side (+40°)and forbids all the movements
This indicates a system failure with the column incidence end position on head
side (-40°)and forbids all the movements
This indicates a system failure with the table tilting end position on foot side
(+90°)and forbids all the movements
This indicates a system failure with the table tilting end position on head side
(-25°) and forbids all the movements
This indicates a system failure with the table variable height in higher position
and forbids all the movements
This indicates a system failure with the table variable height in lower position
and forbids all the movements
This indicates a system failure with the tabletop longitudinal end position on
foot side and forbids all the movements
This indicates a system failure with the tabletop longitudinal end position on
head side and forbids all the movements
This indicates a system failure with the tube movement (SID) in higher position
and forbids all the movements
This indicates a system failure with the tube movement (SID) in lower position
and forbids all the movements
This indicates a system failure with the tabletop lateral end position on front
side and forbids all the movements
This indicates a system failure with the tabletop lateral end position on rear side
and forbids all the movements
This indicates a system failure with the autopositioning and forbids all the
movements
This indicates a system failure with the automatic collimator and forbids all the
movements
Regarding the number of exposures selected for stitching, the carriage position is
too close to end of table.
Regarding incidence and number of exposures selected for tomography, the
carriage position is too close to end of table.
38
Press OK.
- If the message does not
appear again, move the
system slightly in the
other way and try again
the movement.
Then, if the message
appears, turn the system
OFF and call Field
Service.
- If the message appears
again, turn the system
OFF and call Field
Service
Move the carriage up
and try again the
acquisition