 Equipment described herein is unsuitable for operation where and when flammable gasses or vapours
are present.
 It is vital that an authorized distributor carries out the assembly, extensions, adjustments, modifications
and repairs. Also, your radiology unit must be installed in premises that comply with IEC provisions and
the standards in force.
 Any modification, not recommended or not approved by the manufacturer, is prohibited
 Due care must be taken to prevent injury to the patient. Ensure that patient's clothing cannot be trapped
in the equipment. The user must ensure that the patient remains to the specified handgrips.
 For examinations without using the detector inside support, that is to say for direct exposure in the
room, for example with one additional portable detector put on the tabletop (elbow, hand ...), the user
must position the patient at the head of the table and not on lateral side.
 The normal load of patients should be done when the tabletop is in the central position
 The system « 120 cm tabletop movement » works only for patient loading in lying position, from
stretcher to tabletop. In this case, the heights of the stretcher and the table must be identical. Never let
the patient at the end of the tabletop. When the tabletop is out, do not sit or exert pressure on the
tabletop.
 The patient must be correctly and securely positioned (using handgrips, footrest...), before moving or
tilting the table.
 The user must always ensure that there are no obstacles that would interfere with the table. Ensure that
accessories such as shoulder supports, leg supports or footrest are correctly fixed to the table.
 When the table is switched on, an operator must be in the exam room while a patient is on or close to the table.
 In the event of mains supply microcuts, control inverter fault may happen for the motor of carriage,
column, tilting or longitudinal tabletop movement (120cm one). Then you have to switch off the table
with the general ON /OFF and switch it on again two minutes later.
 Safety signs:
User's Manual
General warning sign
Warning: dangerous voltage
Follow operating instructions
CAUTION : X-RAYS
10