1.2 TABLE CONTROL CONSOLE
Note: a seconday console is optional. The way to use it is almost the same as for the main console.
On secondary console, a locking key activates the main console (and thus activates the infrared control
handswitch and inactivates the secondary console) OR activates the secondary console (and thus
inactivates the infrared control handswitch and inactivates the main console).
The controls of axes as collimator shutters, detector rotation (for version with removable detector), or
zoom change can be done on the two consoles whatever locking system position is.
The secondary console allows triggering the emission of X-rays, but it is not possible to perform parameter
settings, Generator interface is not available.
WARNING: The secondary console should only be connected to the power cable located in the table output
1.2.1.
TABLE SWITCHING ON
1 (I) button on the back of the control console switches on the table.
1
1.2.2.
TABLE SWITCHING OFF
(O) button on the back of the control console switches off the table.
(O) button on the back of the control console swicthes off the generator
One emergency stop button on the main control console (n°2)
Two other emergency stop buttons are located on the table, one on each side of detector support.
Note: these emergency stop buttons only stop the table movements.