DDS Calorimeters CAL3K Series Operation Manual page 52

Hide thumbs Also See for CAL3K Series:
Table of Contents

Advertisement

114
READ HELP TEXT
Each fault or operation has a message number attach, the number starting with a 'M' followed by 3
numerical digits. Type the number with an 'M' or without it and the full help text is displayed.
46
READ MAX VESSEL
The vessel maximum data are displayed. The vessel must be in the well.
89
READ PRESSURE
The actual vessel pressure is displayed until ESC. This applies to the CAL3K_AP only.
115
READ PRESSURE REC
The unit records any vessel pressure excess above the PEAK LIMIT (CMD=87) and above the absolute
maximum of 90bar. This command reads the recordings from the memory. Press {enter} for the next
and ESC to terminate reading. The pressure recordings can be cleared by CMD=38 and printed by
CMD=101. A pressure record can be simulated with CMD=126, of course the data are junk.
50
READ MEM RESULT
This command reads results from the memory to the display starting with the last one first. Press ENTER
to display the next result. The result in the display can be printed by pressing F8.
65
READ RESULT DAY:
This command displays the number of results for the entered day number 1-31.
36
READ RISE VESSEL
The temperature rise of the vessel during the main period can be displayed until ESC.
68
READ STABILITY
The vessel stability (2
43
READ TEMP VESSEL
The vessel temperature is displayed until ESC.
83
READ VESS.FIRING
The number of firings of the inserted vessel are displayed.
DDS CALORIMETERS OPERATIONS MANUAL V1.1
0
0
0
0
0
0
0
1
nd
derivative of temperature) can be displayed until ESC.
1
1