This command simulates a result (junk data). If the print output is enabled (CMD=63, SET OUTFORM A-
H) and a printer is connected then the junk data appear on the output.
The following print outputs are possible:
69
PRT RESLTS FORM:
Enter the output form descriptor A-H and all results in the memory are printed. The selected (CMD=63)
form remains as specified.
52
PRT RESULT DAY:
Enter the day number 1-31 and all results recorded at this day are printed. The result memory should be
cleared every month, CMD=39.
64
PRT RESULT GID:
Each result in memory has a GID (Group Identification) of 16 characters. The GID can be an entered text
(CMD=3) or the date (CMD=122). Enter the GID (Up to 16 characters) and results are printed.
57
PRT TODAY FORM:
Enter an output format A-H and all results for today are printed.
67
PRT CV'S VESSEL:
The entry is a 4 character vessel serial number, printed on the vessel lid. The complete result memory is
searched for this serial number and all results are printed.
55
SET FILTER C,F:
A filter can be applied to the print output
The print outs can be filtered to exclude Calibrations (C) and Faults (F).