4.3
COMMAND NUMBER EXECUTION
All the keyboard commands are indexed with a number, which is indicated in this manual with each key
sequence. This number is used to report a problem to the factory (Software bug) when a unit is used
with a foreign keyboard. It bridges any possible language barriers. The command index, Command text
and command priority is indicated. Example: CMD=54
54
D1 PORT SPEED
54
D1 PORT SPEED
0
To execute a command by the number you need to know its Index first. Type 'XX' once to enable
command number execution. No password entry is needed for command number execution. The 'XX'
entry is cleared after the standard password timeout.
5
TESTING
5.1
TEST CAL3K HARDWARE
The test is activated by F9 or by:
13
CAL3K-TEST:
This command tests the CAL3K hardware. Once started the unit will cycle through all tests and will
'hang' on any fault. Individual test can be executed (without cycling through all preceding test) by
replacing the 'YES' entry by the test number.
This test should be repeated once a day. The vessel must not be in the well.
5.2
TEST VESSEL HARDWARE
This test is activated by F7 or:
14
VESSEL-TEST:
The vessel must be in the well.
CAL3K-AP only: The oxygen pressure must be connected. All vessel functions are tested and a summary
is displayed. The test must be repeated for each vessel.
DDS CALORIMETERS OPERATIONS MANUAL V1.1
0
=
Command Number Index
=
Command text
=
Priority, Open, No password required
0
0
2022
Digital Data Systems (Pty) Ltd.
CAL3K Advanced Bomb Calorimeters
www.ddscalorimeters.com
Website:
www.ddscalorimeters.com
7