DDS CALORIMETERS
OPERATIONS MANUAL
VERSION NO: 1.1
DATE: 2022
MANUFACTURED BY: DIGITAL DATA SYSTEMS
!
!

Advertisement

Table of Contents
loading

Summary of Contents for DDS Calorimeters CAL3K Series

  • Page 1 DDS CALORIMETERS OPERATIONS MANUAL VERSION NO: 1.1 DATE: 2022 MANUFACTURED BY: DIGITAL DATA SYSTEMS...
  • Page 2: Table Of Contents

    OUTPUT, EXTERNAL INITIATED ..................17 9.6 OUTPUT TO A SPEADSHEET ..................17 9.7 OUTPUT TO AN EXTERNAL PC ..................17 10 CYCLES ..........................18 10.1 CYCLE: PREPARATION ....................18 10.2 CYCLE: FILLING (CAL3K-AP ONLY) ................. 18 DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 3 FOOD COMPENSATION ....................32 17 LIMITS ..........................33 18 TIME AND DATE ......................... 34 19 3K-AP PRESSURE COMMANDS .................... 35 20 PRINTING ........................... 38 20.1 PRINTING: SETTING THE PORT SPEED ................38 DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 4 32 MAINTENANCE ........................69 32.1 DAILY USE ........................69 32.2 MATERIALS TESTED ...................... 69 32.3 ENVIRONMENT ......................69 32.4 MAINTENANCE SCHEDULE (GUIDELINE) ..............69 33 WARRANTY – FOR NEW PRODUCTS ..................71 DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 5 PURPOSE OF PRODUCT WARRANTY ................71 33.2 OWNER’S WARRANTY ....................71 33.3 EXCLUDED (NON-WARRANTY) PARTS ................71 33.4 WARRANTY – REPAIRED PRODUCTS ................72 33.5 GENERAL – WARRANTY VOID ..................72 33.6 WARRANTY REGISTRATION FORM ................73 DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 6: Introduction

    If the sample explodes then some sample material is settling unburned on the vessel wall (which makes the result useless) and the pressure rises too high which is un-safe. Therefore: DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 7: Safe Electrical Installation

    Should the Oxygen supply be in another room, make sure there is a high pressure shut off valve near the CAL3K calorimeter. Start the oxygen regulator at a very low (Adjustor turned OUT) reading and then adjust it until 30bar = 3000KPA is indicated on the gauge. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 8: Over-Pressure

    A deflating cap 3K-3-22 (manual deflation) and suitable low pressure plastic hose is provided. The CAL3K-AP has a special exhaust port and the exhaust plumbing parts are provided as well. Note: There is no pressure in the exhaust pipe. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 9: Safe Vessel Handling

    NOTE: THE MANUFACTURER (DDS) AND ITS AGENTS CANNOT BE HELD RESPONSIBLE FOR ANY LOSS OR DAMAGE RESULTING FROM NEGLIGENCE OR INCORRECT USE OF THEIR EQUIPMENT. IT IS YOUR DUTY AND RESPONSIBILITY TO CHECK ALL SAFETY ASPECTS OF THE CAL3K SYSTEM REGULARLY. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 10: Basic Keyboard Operations

    The keyboard commands below will be written in a different font, italic and indented. SET OPS MODE The CAL3K has four operational modes: Dynamic Adiabatic Isothermal Fast Dynamic To reset the CAL3K to a working condition you need to install the default parameters by: DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 11: General Keyboard Access

    D) A number display indicating the PRESENT value At this point you can either: Terminate, start fresh Enter a NO, YES, Number, text ENTER Make the change permanent: The NO, YES, Number, text takes affect DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 12: Command Number Execution

    The vessel must be in the well. CAL3K-AP only: The oxygen pressure must be connected. All vessel functions are tested and a summary is displayed. The test must be repeated for each vessel. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 13: Test The Fan

    Note that the filling is very fast because the volume (no =Vessel) is very small. Replace the filling nut again. The filling nut is plastic and must be tighten by hand. Do not use tools and do not ‘cross thread’ it. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 14: Test Vessel Leak

    13. CAL3K-AP: the vessel deflates automatically to the exhaust port and the lid opens. 14. CAL3K-A/F: The lid opens automatically. 15. CAL3K-S/ST: The lid must be opened manually. 16. Remove the vessel for cooling and subsequent new preparation. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 15: Mode Changes

    Each of these modes has its own calibration, which means the vessel(s) must be calibrated in the new mode. Each of the modes has its own default parameters. The most preferred mode is 0=dynamic, followed by 3=Fast Dynamic. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 16: Connecting

    The port speed is settable by: D1 PORT SPEED This entry sets the D1 RS232 port speed in baud. The default is 19200baud. The speed can be set from: 1200 to 115200baud. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 17: Connecting To A Pc With A Rs232 Port

    The balance must have the same setup as the CAL3K. The setting is: Parity None Stop bits Data 8 bit Default Speed 2400 The speed in the CAL3K can be adjusted from 1200 to 38400 baud. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 18: Output

    The formats are not set in stone, so it is best to print the HEADER of all formats. The text descriptions give you are faint idea of what the number columns(data) are. PRT A-H HEADER The unit has 8 print (Output formats) named A-H. The format can be set by SIMULATE RESULT DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 19 SET FILTER C,F: A filter can be applied to the print output The print outs can be filtered to exclude Calibrations (C) and Faults (F). DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 20: Output Printing, Keyboard Initiated, Other

    Exceeding the internal max pressure setting 90bar The pressure recordings can be reset by CLEAR PRESS RCRD, CMD=38. The records can be read out on the display with CMD=115 and a junk record can be simulated with CMD=126. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 21: Output Printing, Internal Initiated

    The capturing of the data can be done by any terminal emulator like TERA TERM VT. Connect to the PC with a RS232 cable, then configure it for the RS232 baud speed (CMD=54). Of course, you can capture DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 22: Output, External Initiated

    __HI;a;l Upload results form ‘A-H’ from record ‘a’, length ‘l’ __HH Upload form ‘U’ header __HK Upload form ‘N’ header __HJ Upload form ‘A-H’ header __HS;n Upload results day ‘n’ form ‘U’ DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 23: Cycles

    This is a very important cycle: it allows the vessel temperature to settle down and equalise. The un- equalized temperatures are the result of filling and handling of the vessel. Vessel communication M210 Initial cycle interruption M216 CAL3K-AP: Filling Leak M304 Vessel temperature M258 DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 24: Cycle: Main

    At power-up the vital parameters are checked during operation, and if they are out of specification then the DEFAULT PARAMETERS must be loaded. PARA DEFAULT LOAD DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 25: Entering Password

    ‘ERASE’ the mass or ‘UNLOCK’ it is as follows: 1. Type the word ‘MASS’, Enter, a beep can be heard and the entered mass is displayed. 2. Press the ‘DELETE’ button DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 26: Sample Identification

    It is useful to test the firing/ignition procedure. It is not recommended for use during normal operation, as this will affect the accuracy of your sample being determined. FIRE MANUALY DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 27: Read Result

    The unit works in KJ/g and Centigrade. However, the CV value can be expressed in other units of measure by: UNIT: KJ,KCA,KBU This command sets the measuring units. Note that the internal working are in KJ/g, but any output is converted to the units. Enter: KJ/g, MJ/Kg KCAL/g KBTU/lb DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 28: Calibration

    It is therefore imperative that the calibrations are performed at different operating conditions like: First/last time in a day During routine operation Whenever the operating conditions have changed Whenever parts (consumables) of the vessel have changed DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 29: Calibration In Different Modes

    The result (and calibration) quality is judged by the compensation amount in Degree C. See: COMPENSATIONS. The compensations are judged by: A) Total compensation. The sum of all compensations: Total Compensation B) The ABSOLUTE sum of all compensations: Maximum Compensation DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 30: Average Calibration

    A ‘ZERO’ entry (Default) disables the acceptance check. A practical example: Average calibration: 2 Average acceptance: 0.024 Then the average will be performed with THREE calibrations if all are excellent to go. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 31: Calibration Summary

    The ‘Spike’ must be of a known calorific value, usually benzoic acid of approx. 26.454MJ/Kg. On some occasions a Gelatine Capsule (3k-4-068) or Crucible Cover Disc (CAL2K-FLDS) needs to be used as a Spike. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 32: Spike Value

    This is the spike mass. If the value is >0 the spiking is assumed and the result will be reduced by: Spike mass x Spike value. The spike mass is cleared after a determination. A ‘ZERO’ spike mass disables the spiking process. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 33: Step By Step Entering The Spike Mass

    Of course, you can determine the average CV of let say 10 capsules. The Temperature Rise of each determination must be read out via the keyboard (CMD=36) and then averaged and entered via CMD=29. The EASY SPIKING is enabled via: GELATINE CAPSULE DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 34: Corrections

    If the sample contains sulphur then energy is released during the formation of SO2 from oxygen and sulphur. This energy is deducted from the nett CV. The sulphur content of the sample must be measured before and the CV (KJ/g) of SO2 formation must be entered: DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 35: Moisture Correction

    Then the temperature rise caused by a gelatine capsule is entered here (Default=2.594). This entry is associated with the capsule (Manufacturer, weight, size) and is constant unless changed. If a capsule is used (for powdery samples) then is must be enabled with CMD=28. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 36: Firing

    These compensation factors change with: Vessel construction: Deflector plate, Crucible holder, firing wire length, cotton length CAL3K construction: Ambient temperature, ATC temperature (if installed), Ventilation Vessel age and surface condition DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 37: Moisture Compensation

    However, a calorimeter is a nice, fast and convenient tool to determine the calories. Food is measured in CALORIES or KJ. The ATWATER specification reads: Protein 4Kcal 16.736KJ Carbohydrates 4Kcal 16.736KJ 9Kcal 37.656KJ Alcohol 7Kcal 29.288KJ DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 38: Limits

    A limit can be set up to warn the operator that the sample mass was too low. The default setting is 8C. A lower rise can be achieved by calibrating the unit to an 8C+-2C. It is just difficult to split a Benzoic Acid tablet to 2/3. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 39: Time And Date

    The time can be used as a Sample Identification (SID). This is not available on the CAL3K-S. When this feature is enabled then the auto-increment of the SID is disabled. TIME IS SID DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 40: Ap Pressure Commands

    The actual pressure may be 1bar down because of pressure settling. The target should be 30bar. Lower filling pressures are indicated for powdery substances to reduce scattering of sample material. The default is 30bar. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 41 If a limit is exceeded the vessel is deflated to a safe pressure and the main period is terminated after a short time. The CAL3K-AP records: The absolute pressure The time from firing Peak Temperature, approx. 30 seconds after firing DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 42 The test pressure can be set by: PRESS LEAK TEST This command sets the leak test pressure applied during SYSTEM LEAK TEST, CMD=94. The leak test is useful when a leak is suspected. The default value is 25bar. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 43: Printing

    It is possible that some output forms change in future. It is therefore advisable that you check the output forms A-H for suitable data in a printable form. Use: PRT TODAY FORM: DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 44: Printing A Form Header

    Note that the FULL GID is required. If GID = Date is set then all results of a particular date are printed. The print format may not show the GID or not all of it. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 45: Printing Enable

    A printout of the CAL3K configuration and setup can be obtained by: PRT CALORI INFO This printout shows all the relevant information of the CAL3K. 20.9.2 PRINT VESSEL INFO This printout shows all information of the inserted vessel: PRT VESSEL INFO DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 46: Print Vessel Calibration

    3. Start the App and configure it to the same speed as the CAL3K port 4. If you have no results in memory then simulate a result: SIMULATE RESULT 5. Type the print command on the CAL3K keyboard: PRT TODAY FORM: DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 47: Printing Real Time

    The letter ‘C’ enables the printing of the customer header information and the letter ‘P’ the printing of the relevant parameters before the real time printing starts 30 seconds before firing. Optional disable result printing: PRT ENABLE RESULT Enter ‘NO’. This switches the result printing off. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 48: Print Result Dumps

    The output data are therefore devoid of any computation, adjustments or compensation. In addition, the output protocol can be preceded by a: USER HEADER: Name of institution, faculty, division PARAMETER HEADER: The list of parameters under which this experiment was performed DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 49: Clearing

    25.1 CLEAR RESULT MEMORY CLEAR RESULTS This command clears all results. Depending on the unit the CAL3K keeps from 300 to 700 results. Enter ‘YES’ to clear the result memory. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 50: Clear Help Events

    Firing count at max Temperature Date and time of max Temperature The vessel must be inserted for the clearing. The command should be repeated with each vessel. 25.6 CLEAR PRESSURE RECORD CLEAR PRESS RCRD DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 51: Reading

    The number of firings in this chassis are displayed. READ DRIFT The drift (Temperature difference per 6 seconds) is displayed until ESC. READ FIRING VOLT The actual firing voltage is displayed until ESC is pressed. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 52 The vessel stability (2 derivative of temperature) can be displayed until ESC. READ TEMP VESSEL The vessel temperature is displayed until ESC. READ VESS.FIRING The number of firings of the inserted vessel are displayed. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 53: Bluetooth

    This is a service command. It allows checking out the complete calorific determination but 6 times faster without oxygen or balance. See: DEMO (CMD=108) for the slowed down version. Note: The simulated result is entered in to memory, but is suitable flagged. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 54: Simulating Calibration Process

    The calibration entry can be displayed with: READ BOMB CALI The bomb calibration field (all entries) are cleared with: CLEAR B_CAL S/N DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 55: Simulating A Help Event

    This command is for demonstration only. It allows to demonstrate the complete operating cycle without: Oxygen pressure Sample insertion Mass entry It is possible to perform a DEMO CALIBRATION. If a demo calibration needs to be done then set calibration active: DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 56: Simulating An Over-Pressure Record

    SET RTHEAD EN CP 1 This command applies to the CAL3K-AP/A/F units with real time (RT) printing. Enter ‘C’ for customer header and ‘P’ for parameter header. See: REAL TIME OUTPUT for details. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 57 This command applies to the CAL3K-A/A only. Enter the numbers 0-3 to select an operating mode. See: OPERATIONS for details. SET RESULT ENABL 1 Enter ‘YES’ to enable the printing of spontaneous result messages. See: PRINTING for details. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 58: All Cal3K Commands

    This is the spike mass. If the value is >0 the spiking is assumed and the result will be reduced by: Spike mass x Spike value. The spike mass is cleared after a determination. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 59 The test must be repeated for each vessel. SPIKE VALUE The Spike Value is the CV of the spike material. If Benzoic Acid is used for spiking, then the CV=26.454KJ/g. The default setting is 26.454KJ/g DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 60 KJ/g, MJ/Kg KCAL/g KBTU/lb Enter the letters: for KJ/g for KCAL/g for KBTU/lb The following conversion applies: 1KJ/g 429.9226 BTU/lb 0.4299226 KBTU/lb 1KJ/g 0.2388459 KCAL/g Therefore: The 26.454KJ/g 11.373 KBTU/lb 6.3184 KCAL/g DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 61 (CMD=28, NO) In the ‘EASY SPIKE’ method the Capsule Rise value (CMD=29) is deducted from the measured temperature rise (Sample + Capsule) before the result is calculated. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 62 Pressure Peak pressure time (Milliseconds) Note that the pressure recordings are triggered by exceeding the ‘PRESSURE PEAK LIMIT’ CMD=87. The recordings are in a circular fashion and the oldest one is over-written. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 63 The vessel maximum data are displayed. The vessel must be in the well. READ CAL3K MAX The CAL3K maximum data are displayed. READ VESSEL INFO The information of the inserted vessel is displayed. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 64 This command prints the vessel information. The vessel must be in the well. SET RTHEAD EN CP The REAL TIME output (from 30 seconds before firing) can be preceded by Customer (C) and Parameters (P). If none is desired enter anything but CP. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 65 The unit can be switched to high display result resolution by a ‘YES’ entry. The resolution is: HIGH XX.XXXX XX.XX Note that the resolution switch affect the DISPLAY only. All digital outputs depend on output format and destination. An output to a PC has a resolution of: XX.XXXXX DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 66 OUTPUT TO A SPREADSHEET. INITIAL TIME The initial time can be set in the CAL3K-A/AP in the experimental mode. READ LAST N_RISE 1 The last nett rise is displayed after the main period. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 67 The actual pressure may be 1bar down because of pressure settling. The target should be 30bar. Lower filling pressures are indicated for powdery substances to reduce scattering of sample material. The default is 30bar. READ PRESSURE CAL3K-AP command. The actual vessel pressure is displayed until ESC. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 68 The reason for slow filling could be: Empty oxygen bottle or blocked inlet filter. The filter (and filling restrictor) are a very unlikely reason for slow filling and the oxygen pressure should be investigated first. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 69 The result is affected by the temperature rise after firing. This affect is minimised by calibrating the unit for high rise (e.g., 0.75g Benzoic Acid). However, the sample mass should be adjusted so that the temperature rise is near to the optimal 12C +-2C. The default setting is 16C. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 70 Each fault or operation has a message number attach, the number starting with a ‘M’ followed by 3 numerical digits. If it is required to read the help text type the command and number. Example: 114, enter, 524 DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 71 Enter a 4-digit (or less) serial number and a calibration record is entered in to the specified memory (and mode). This calibration is useless together with valid calibrations and must be cleared before use. The vessel doesn’t have to be inserted, but the serial number must be entered. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 72: Temperature Rise

    The following parameters must be set to confirm with the changed calibration: LOW RISE LIMIT The default is 8C. Change to suit! HIGH RISE LIMIT The default is 16C. Change to suit! MIN MASS LIMIT The default is 0.2g. Change to suit! DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 73: How To Measure The Wire+Cotton Rise

    READ LAST N_RISE 0 This reads the Net Rise of the last sample. This is the temperature rise produced by the gelatin capsule. Enter this value in: 29 CAPSULE RISE C DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 74: Maintenance

    CAL3K Advanced Bomb Calorimeters www.ddscalorimeters.com 32 MAINTENANCE The DDS Calorimeters are designed to work in a laboratory. However, all laboratories are different and the maintenance depends largely on: 32.1 DAILY USE This works out to samples per day. The maintenance of the system is fairly in line with the usage of the system.
  • Page 75 When bottom distorts and becomes completely discoloured DEFLECTOR PLATE (3K-4-005): 30-150 samples Weekly to Monthly, or when centre electrode O-ring is changed. Wipe top of deflector plate clean weekly. FILLING STATION NOZZLE O-RING (3K-3-29): 50-100 samples DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...
  • Page 76: Warranty - For New Products

    User. The manufacturer is not responsible for any damages, consequential, indirect or direct, or loss whatsoever, resulting out of electrical shock, Vessel explosion, or any other cause. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 77: Warranty - Repaired Products

    Warranty will also be void if any of the branding stickers or decals have been removed and/or the colour of the Scientific product has been changed in any way. DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website:...
  • Page 78: Warranty Registration Form

    NAME (printed): ____________________________ WITNESS: ____________________________ I have read, understood and accepted the conditions of this warranty, the details of which are printed in this Manual. CUSTOMER’S SIGNATURE: __________________________________________________ NAME (printed): __________________________________________________ DATE: ___________________________________________ DDS CALORIMETERS OPERATIONS MANUAL V1.1 2022 Website: www.ddscalorimeters.com...

This manual is also suitable for:

Cal3k-apCal3k-aCal3k-fCal3k-stCal3k-s

Table of Contents