Section 3: Adjustment
10 mA range (-) full scale nominal
cal_DCI_10mA_fs_neg
cal_DCI_DIGI_10mA_fs_neg
*Characterized step measured by the Fluke 8508A or 8588A
Command parameters
cal_DCI_100mA_fs_pos
cal_DCI_DIGI_100mA_fs_po
s
cal_DCI_100mA_fs_neg
cal_DCI_DIGI_100mA_fs_ne
g
cal_TS5
cal_dcI_1A_fs_pos
cal_DCI_DIGI_1A_fs_pos
cal_DCI_1A_fs_neg
cal_DCI_DIGI_1A_fs_neg
cal_DCI_DIGI_3A_fs_pos
cal_DCI_DIGI_3A_fs_neg
DC current 10 A range adjustment
The following procedure provides instructions for completing a rear-terminal dc current adjustment for
the 10 A range. The following section provides a command parameter table to complete the
adjustment.
For this adjustment, you need a:
•
Fluke 5720 calibrator
•
10 A cable
3-24
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
dc current 0.022
-0.01
dc current 0.022
-0.01
Calibrator
Calibrator range Calibrator
function
dc current 0.22
dc current 0.22
dc current 0.22
dc current 0.22
dc current n/a
dc current 2.2
dc current 2.2
dc current 2.2
dc current 2.2
dc current 2.2
dc current 2.2
8508A or 8588A
10 mA
Reading
8508A or 8588A
10 mA
Reading
Calibrator
Reference
value
stimulus
DMM dc
value
current range
0.1
n/a
20 A
0.1
n/a
20 A
-0.1
n/a
20 A
-0.1
n/a
20 A
n/a
n/a
20 A
1
n/a
20 A
1
n/a
20 A
-1
n/a
20 A
-1
n/a
20 A
2
n/a
20 A
-2
n/a
20 A
DMM6500-905-01 Rev. E June 2021