Section 3: Adjustment
Command parameters for an ac current adjustment
When calibrating your DMM6500 for an ac current adjustment, use the following command
parameters.
Send each command parameter twice. First, send the parameter using the setup command. Second,
send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_ACI_100uA_1kHz_tenth")
cal.adjust.step.execute("cal_ACI_100uA_1kHz_tenth")
Command parameters
cal_ACI_100uA_1kHz_tenth
cal_ACI_100uA_1kHz_fs
cal_ACI_1mA_1kHz_tenth
cal_ACI_1mA_1kHz_fs
cal_ACI_10mA_1kHz_tenth
cal_ACI_10mA_10Hz_fs
cal_ACI_10mA_10kHz_fs
cal_ACI_10mA_1kHz_fs
cal_ACI_100mA_1kHz_tenth
cal_ACI_100mA_1kHz_fs
cal_ACI_100mA_10Hz_fs
cal_ACI_200mA_1kHz_fs
cal_ACI_200mA_10kHz_fs
cal_ACI_1A_1kHz_tenth
cal_ACI_1A_10Hz_fs
cal_ACI_1A_1kHz_fs
cal_ACI_3A_1kHz_tenth
cal_ACI_3A_1kHz_fs
3-30
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Calibrator
Calibrator
function
range
ac current
0.00022
ac current
0.00022
ac current
0.00022
ac current
0.0022
ac current
0.0022
ac current
0.022
ac current
0.022
ac current
0.022
ac current
0.022
ac current
0.22
ac current
0.22
ac current
0.22
ac current
0.22
ac current
0.22
ac current
2.2
ac current
2.2
ac current
2.2
ac current
2.2
Calibrator
Calibrator
value
frequency
0.00001
1 kHz
0.0001
1 kHz
0.0001
1 kHz
0.001
1 kHz
0.001
1 kHz
0.01
10 Hz
0.01
10 kHz
0.01
1 kHz
0.01
1 kHz
0.1
1 kHz
0.1
10 Hz
0.2
1 kHz
0.2
10 kHz
0.1
1 kHz
1
10 Hz
1
1 kHz
0.3
1 kHz
2
1 kHz
DMM6500-905-01 Rev. E June 2021