DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Range
Nominal
input (µA)
1 mA
1
0.5
–0.5
–1
DC current 100 mA to 3 A range verification
To verify dc current accuracy on the 100 mA to 3 A ranges, you will:
Apply accurate current from the dc current calibrator directly to the DMM6500 front-panel
terminals.
Verify that the displayed readings are within specified limits.
To verify dc current accuracy:
1. Set up the DMM6500 for dc current and the range being tested. Make sure that relative offset is
disabled.
2. Connect the DMM6500 and calibrator as shown in the following figure.
DMM6500-905-01 Rev. E June 2021
Specification
tolerance (µA) (based
on nominal)
0.000500
0.000275
0.000275
0.000500
Figure 12: Connections for 100 mA to 3 A range verification
Section 2: Performance verification
Range
Nominal
Specification
input (µA)
tolerance (µA)
(based on nominal)
10 mA
10
0.00250
5
0.00150
–5
0.00150
–10
0.00250
2-33