DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual

cal.adjust.step.execute()

This function executes the specified adjustment step.
Type
TSP-Link accessible
Function
Yes
Usage
cal.adjust.step.execute(stepname)
cal.adjust.step.execute(stepname, value)
stepname
value
Details
This command generates an error if:
Calibration is locked
The step does not complete successfully
The value that is passed is invalid for the step, out of range, or not needed
Also see
cal.lock()
(on page 4-6)
cal.unlock()
(on page 4-9)
DMM6500-905-01 Rev. E June 2021
Affected by
The adjustment step to start
Value for this adjustment step. If using the default then value is optional
Section 4: TSP command reference
Where saved
Default value
4-5