Keithley DMM6500 Manual page 32

6 1/2-digit multimeter with scanning
Hide thumbs Also See for DMM6500:
Table of Contents

Advertisement

Section 2: Performance verification
Equation for 0 °C to 850 °C temperature range
R
= R
(1 + AT + BT
RTD
0
where:
R
is the calculated resistance of the RTD
RTD
R
is the known RTD resistance at 0 °C
0
T is the temperature in °C
A = alpha [1 + (delta/100)]
B = –1 (alpha)(delta)(1E-4)
The alpha and delta values are listed in the following table.
RTD parameters for equations
The RTD parameters for the Callendar-Van Dusen equations are listed in the following table.
DMM6500 resistance to temperature device (RTD)
Type
Standard
PT100
ITS-90
D100
F100
PT385
IPTS-68
PT3916
2-22
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
2
)
Alpha
Beta
0.00385055
0.10863
0.003920
0.10630
0.003900
0.11000
0.003850
0.11100
0.003916
0.11600
at 0 °C (Ω)
Delta
R
0
1.49990
100.0000
1.49710
1.49589
1.50700
1.50594
DMM6500-905-01 Rev. E June 2021