8. What If ...
1912
yes
1920
yes
2754
yes
2755
yes
2756
yes
2757
yes
2758
yes
2759
yes
2760
yes
**
2761
yes
**
2762
yes
**
*
Furnace head opens if this error occurs!
** A program in progress is aborted
*** Error cannot be acknowledged; programs cannot be started.

8.2 Additional error messages

Please contact the Ivoclar Vivadent After Sales Service, if one of the following error numbers is being displayed:
25, 29, 43, 44, 45, 46, 47, 48, 56
103, 107, 108, 109, 143, 144, 145, 146, 147
148, 700, 701, 703, 704, 707, 1010, 1011, 1012
1013, 1014, 1015, 1016, 1017, 1018, 1019, 1024, 1025
1026, 1028, 1143, 1144, 1145, 1146, 1147, 1148, 1207
1300, 1301, 1303, 1304, 1305, 1401, 1402, 1500, 1750
1751, 1752, 1753, 2750, 2751, 2752, 2753, 2770
72
Data backup – recover
Data backup – recover
Error self-test
Error self-test
Error self-test
Error self-test
Error self-test
Error self-test
Error IRT
Error IRT
Error IRT
The new software includes program groups which were not yet available at
the time of the data backup. Please check the individual program groups.
The individual program groups are no longer valid.
The data backup could not be performed. Please install a newer software
version.
An error occurred while checking the infrared camera
An error occurred while checking the infrared camera
An error occurred while checking the infrared camera
An error occurred while checking the infrared camera
An error occurred while checking the infrared camera
An error occurred while checking the infrared camera
Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. Check if the protective foil has been removed from the sensor or if
the sensor has been contaminated. Check if a correct firing tray is used.
The height of the firing tray must correspond with the height of the firing
trays recommended for the respective materials by Ivoclar Vivadent AG.
Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. The object was not sufficiently dried in the specified time. Please
repeat the procedure after the object has cooled down.
An error occurred while trying to identify the object. Check if interfering
heat sources are in the visual range of the infrared camera. Sources of
interference are for example: other furnaces, air conditioning systems,
heaters, light bulbs, Bunsen burners, etc. Do not move the objects after the
start of a program.