Error Messages; Safety, Care, Accuracy Test And Disposal - Microlife BP A3 Plus Manual

Hide thumbs Also See for BP A3 Plus:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Only use the Microlife mains adapter available as an orig-
inal accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket 5 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.

9. Error messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «Err 3», is displayed.
Error
Description Potential cause and remedy
«Err 1»
Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat
the measurement.*
«Err 2»
Error signal
During the measurement, error signals
were detected by the cuff, caused for
BN
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«Err 3»
No pressure
An adequate pressure cannot be
in the cuff
generated in the cuff. A leak may have
BK
occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
«Err 5»
Abnormal
The measuring signals are inaccurate
result
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.*
«Err 6»
MAM Mode
There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for performing
reliable measurements and then
repeat the measurement.*
6
Error
Description Potential cause and remedy
«HI»
Pulse or cuff
The pressure in the cuff is too high
pressure too
(over 299 mmHg) OR the pulse is too
high
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
«LO»
Pulse too low The pulse is too low (less than 40 beats
per minute). Repeat the measure-
ment.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.

10. Safety, care, accuracy test and disposal

m
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical specifications» section.
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient's symptoms. Do not
rely on the measurement result only, always consider other
potentially occurring symptoms and the patient's feedback.
Calling a doctor or an ambulance is advised if needed.

Advertisement

Table of Contents
loading

Table of Contents