Emerson Rosemount 3812 Maintenance And Troubleshooting Manual page 29

Liquid ultrasonic flow meters for direct or remote mount meter electronics
Hide thumbs Also See for Rosemount 3812:
Table of Contents

Advertisement

Maintenance and Troubleshooting Manual
00809-0100-3812
Table 2-2: Maintenance
MeterLink utility
MeterLink Logs/ Reports Menu Meter performed a Warm start
MeterLink Tools → Edit/
Compare Configuration
Menu
MeterLink Logs/Reports Menu Power failure
Rosemount 3812 Liquid Ultrasonic Flow Meter
(continued)
Diagnostics
or a warm start is required
Meter performed a Cold Start
Troubleshooting
April 2022
Action(s)
Meter performed a Warm Start:
Collect an Archive event log (Audit log)
using MeterLink to view configuration
parameters changes and when they
changed.
Warm start required:
When you make changes to the
transducer characteristics, sample rates,
the device number, or a Modbus map
file.
The meter configuration has reset to
default values and the meter is not
configured correctly to measure flow.
Unless the cold start occurred after
upgrading firmware, you may need to
replace the CPU Module.
If the Cold Start occurred after a firmware
upgrade, you must reconfigure the meter
from a previously saved configuration file
using the Edit/Compare Configuration
screen. Then clear the latched alarm on the
Status Summary page.
If this was a known power fail or restart of
the meter just acknowledge this alarm on
the Status Summary page.
If this was an unexpected restart of the
meter, verify the integrity of the power to
the meter and make sure that the voltage
level is in the range of 11-36 VDC at the
meter.
Collect an Archive event log (Audit log)
using MeterLink.
29